Accuracy
MODEL "H20/PSI
0 to 830 "H2O: ±(0.05% of reading + 0.1"H2O)
0 to 30 PSI: ±(0.05% of reading + 0.01 PSI)
30 to 50 PSI: ±(0.25% of reading)
0 to -14.7 PSI: ±(0.5% of reading + 0.02 PSI)
Over Pressure: 100PSI
MODEL "Hg/PSI
0 to 30 "Hg: ±(0.1% of reading + 0.01"Hg)
0 to -30 "Hg: ±(0.1% of reading + 0.01 "Hg)
0 to 500 PSI: ±(0.1% of reading + 0.02 PSI)
0 to -14.7 PSI: ±(0.25% of reading)
Over Pressure: 1000 PSI
MODEL "HgA/mmHgA
0 to 760 mmHgA: ±(0.08% of reading + 0.2 mmHg)
0 to 30 "HgA: ±(0.08% of reading + 0.01"Hg)
30 to 60 "HgA: ±(0.25% of reading)
Over Pressure: 100 PSIA
MODEL mmHgA/PSIA
0 to 760 mmHgA: ±(0.08% of reading + 0.2 mmHg)
0 to 30 PSIA: ±(0.08% of reading + 0.01 PSI)
Over Pressure: 100 PSIA
Temperature
Operating & Compensated: -10°C to 50°C (14°F to 122°F).
Storage: -40°C to 75°C (-40°F to 167°F).
|
Connections
Pressure connection: 2 x 1/8" NPT (Female)
Banana jacks: 4mm plugs on 19mm (0.75") centers
Media Compatibility
Liquids and gases compatible with PTFE
(Polytetrafluoroethylene) penetrated, hard anodized, aluminum (transducer housing), 316 stainless steel, buna-n (O-ring).
Power
Battery: One standard 9V battery.
Battery life: 400 hours typical (alkaline battery).
Enclosure
Dimensions: 50 mm (2.0â€) W x 117 mm (4.625") L
Weight: 85 g (3 oz.)
Sensor